Boston Scientific Simulates Release Mechanisms in Drug-Eluting Stents
T. Schauer, I. Guler
Boston Scientific Corporation, MN, USA
Stent insertion through the coronary artery is a common procedure used to treat restricted blood flow to the heart caused by stenosis. Following the procedure, restenosis may occur due to excessive tissue growth around the stent. Researchers at Boston Scientific are using multiphysics simulation to better understand how drug-eluting stents prevent tissue regrowth and further improve their design.
In COMSOL Multiphysics, the researchers developed a pore-shell model of a stent coating designed for controlled release of medication into surrounding tissues. In their model, the stiff-spring method was implemented to ensure a continuous diffusive flux of medicine at the pore-shell interface. The Optimization Module was used to refine model parameters that could not be determined experimentally, such as the polymer shell thickness and the retardation coefficient of the pores in the stent coating. Both simulation and experimental data confirm a fast release mechanism from pores and slow diffusion through the polymer encapsulant in the shell.
Suggested Stories
Capacitively Coupled Plasma Analysis
AltaSim Technologies provides engineering consulting services involving advanced multiphysics modeling such as ...
Man-Made Stars: Evaluating Structural Integrity in High-Performance Nuclear Fusion Machines for Power Generation
For over 50 years, power generation through nuclear fusion has been the focus of international research efforts to ...
Patterning Cells with the Flip of a Switch for Bioengineering Applications
Dielectrophoresis (DEP) is a method of patterning cells and particles used in medical, electronics, and energy ...
Reduced-Weight Reaction Sphere Makes Way for Extra Satellite Payload
CSEM is a private applied research and development center specializing in micro-technology, system engineering, micro ...