SIMULATION SUCCESS STORIES

Boston Scientific Simulates Release Mechanisms in Drug-Eluting Stents

Healthcare

T. Schauer, I. Guler
Boston Scientific Corporation, MN, USA

Stent insertion through the coronary artery is a common procedure used to treat restricted blood flow to the heart caused by stenosis. Following the procedure, restenosis may occur due to excessive tissue growth around the stent. Researchers at Boston Scientific are using multiphysics simulation to better understand how drug-eluting stents prevent tissue regrowth and further improve their design.
In COMSOL Multiphysics, the researchers developed a pore-shell model of a stent coating designed for controlled release of medication into surrounding tissues. In their model, the stiff-spring method was implemented to ensure a continuous diffusive flux of medicine at the pore-shell interface. The Optimization Module was used to refine model parameters that could not be determined experimentally, such as the polymer shell thickness and the retardation coefficient of the pores in the stent coating. Both simulation and experimental data confirm a fast release mechanism from pores and slow diffusion through the polymer encapsulant in the shell.

  •  /
1
Loading PDF...

Suggested Stories

Defending Automotive Components Against Corrosive Destruction

Defending Automotive Components Against Corrosive Destruction

Galvanic corrosion, which occurs when the surfaces of different adjoining metals begin to degrade, costs the ...

Complex Fenestration System for Energy Efficiency

Complex Fenestration System for Energy Efficiency

Fenestration systems can greatly impact the energy efficiency of buildings. Eurac Research is using simulation to ...

Designing the Gold Standard of Immersive Audio for Virtual Reality Gaming

Designing the Gold Standard of Immersive Audio for Virtual Reality Gaming

Tectonic Audio Labs used electromagnetics, mechanical, and acoustics simulation to develop a state-of-the-art balanced ...

A 100-Fold Improvement in Lithography Resolution Realized with a 150-Year ...

A 100-Fold Improvement in Lithography Resolution Realized with a 150-Year-Old “Perfect Imaging” System

The idea of "perfect imaging", that is a resolution less than the wavelength of light producing the image, was first ...